Part 1 Principles
1. Fluorescence microscope
2. Filterset in FL-Mic
3. How concocal differs?
4
. What is confocal?
5. Resolution in confocal
6. Optical sectioning
7. Confocal image formation
    and time resolution
8. SNR in confocal
9. Variations of confocal
      microscope

10. Special features from
     Leica sp2 confocal

Part 2 Application
1. Introduction
2. Tomographic view
    (Microscopical CT)

3. Three-D reconstruction
4. Thick specimen
5. Physiological study
6.
Fluorescence detecting
       General consideration
      
Multi-channel detecting
       Background  correction
       Cross-talk correction
            Cross excitation
            Cross emission
            Unwanted FRET


Part 3 Operation and
             Optimization

 1. Getting started
 2. Settings in detail
 
     Laser line selection
      Laser intensity and 
         AOTF control

      Beam splitter
      PMT gain and offset 
      Scan speed
      Scan format, Zoom
        and Resolution

     Frame average, and
         Frame accumulation
     Pinhole and Z-resolution
     Emission collecting rang
        and Sequential scan


When Do you need confocal?
FAQ
Are you abusing confocal?

Confocal Microscopy tutorial

Part 3 operation, optimization of Leica SP2 LSCM

Scan format, Zoom factor and resolution

Scan format,  as called in Leica confocal software LCS, expressed as 512 x 512, 1024 x1024, up to 4096 x 4096, first number for pixels per line, second number for line per frame scanned, is actually an expression of digital image resolution (it is called image resolution in Biorad and Zeiss confocal microscope).

In digital image field, resolution usually takes the form of "pixels x pixels". The real resolving power of the image, in terms of the smallest detail it reveals, depends on size of the final output. That is, on how big area these pixels are distributed, which determine the size of a pixel thus resolving power of the image during image capture process. The resolving power is finally restricted by the resolution of output device: the dpi value of a printer or display screen. This factor usually is not a concern because there is no sampling anymore, just image-to-screen pixel-to-pixel projection. As long as there are enough sampled pixels in the data acquisition to meet resolution of the objective lens (see details below) and the pixel number is enough to feed the display with its 96 dpi screen resolution, or match the printer with some 72-100 dpi output resolution for a given printout size, there will be no pixelation on the output image.

When producing a digital image, it is often unnecessary to know the exact resolving power, but the total pixel number and its distribution format (digital resolution) must be specified before a digital image file can be produced. This number in turn directly influence the size of an output at certain dpi value before the image get pixelated. For example, a file with 512 x 512 pixels can produce an output up to 1.7 x 1.7 inch at 300 dpi without pixelation, while a 1024 x 1024 file produces up to 3.4 x 3.4 inch output at 300 dpi. The convenience to estimate final output size is the major advantage of using digital image resolution.

In digital microscopy field, it is more complicated. The match between digital resolution and optical resolution of an objective, the match between digital resolution and final output size have to be considered at the same time when you choose a digital image resolution for an image file. On microscope side, choose adequate magnification to make the details of interesting bigger than dot size of the recording device.

Match optical resolution of objectives

Choosing of scan format in confocal system affects function of both digitizer (sampling along a line at the chosen number) and scanner (how many lines to be scanned).

To match digital resolution of an image frame with optical resolution of an objective, one needs considering: how many sampling points need to be distributed in the intermediate image field of the microscope to achieve a pixel size that will not compromise the resolving power of the lens. According to Nyquist sampling theorem, at least 2-3 sampling rate has to be used to resolve an underlying structure, i.e., 16 to 36 pixel sampling per Airy disc of a objective lens is needed to guarantee its optical resolution being represented in the acquired digital image.

Taking an example by using 40 x objective of 1.25 NA, at 488 nm excitation:
First, from the optical resolution of the objective: Res=0.51*488/1.25=199 nm
If Nyquist sampling frequency of 2 is used, 199/2=99.5 nm; If sampling frequency of 1.5 used, 199/1.5=133 nm (it is said that confocal can use  sampling rate lower than 2 because of its low background). So the required pixel size is about 100-133 nm.

Then consider the digital resolution of the image you acquired and its pixel distribution in the field of intermediate image plane of the microscope:
For Leica IRE2 of FN 22,  diameter 22 mm, an inner square in the circle is diagonally 22 mm  and 15.7 x 15.7 mm by side.
For a 40 x objective, using scan format of 512 x 512: the pixel size is 700 nm. It does not meet the minimum requirement of the Nyquist sampling. If a 5-7x zoom is used, the pixel size reduced to 100-140 nm, that meets the resolution of the 40x objective at 1.5 to 2 Nyquist frequency (see table 2 below for scan format required by each lens for matching its resolution).

Table-2 shows that 4096 format produces very small pixel size that does not require any scanner zoom. Is it a perfect solution that you can routinely use? No, unfortunately, this format makes pixel dwelling time extremely short. Taking 400 Hz scan speed as example, one line uses 1/400 s that is 2.5 ms, each line is divided by 4096 that makes each pixel-dwelling time down to 0.6 ns. For such a short signal collecting time, the signal is extremely weak, the SNR and image quality become very bad without compensation by slow scan speed and high number of average. At 200 Hz scan speed, a 4096-line frame need 20 seconds per frame, and 8 times or more average makes the total scan time up to 2-3 min, then bleaching become a problem, the high optical resolution you expected by using high digital resolution lost in the noisy, bleached image.

So, the optical resolution is not the only thing you need to consider. The question then become: do you need that 200 nm resolution? or does your specimen provide that small details for the confocal system to resolve? If not, you can take it easy: use lower format plus some zoom factor to get a good image that shows everything you can see from microscope. Or use reduced area by scanner zoom. Table 2 tells us, that by using 1024x1024 scan format, a moderate zoom factor 2-3 meets all objective lens's resolution and is recommended for most applications.

Scanner zoom factor vs. digital zoom

Digital zoom simply blown up an image to a bigger area with its original pixel numbers, that soon reach the limit of screen dpi value and image become pixelated. Some image edit software provide pixel interpolation which avoid pixelation when enlarging an image, but image quality deteriorate with massive addition of pixels. In both way,  image resolution do not become better.

Confocal scanner zoom is a totally different story. The zoom-in function reduces the scan angle that in turn reduces the scan area. With a smaller area scanned (enlarged) using the same pixel number, the pixel size reduced. So, confocal zoom enlarges the image at the same time improves the resolution.

Meet image size requirement of the final output

If the resolution is not your main concern, if you are satisfied with the details presented on  the screen already, then choosing scan format become a question of what is the size of the final output.
For example, a 512 x 512 image file can produce a 1.7 x 1.7 inch ( 4 x 4 cm) print-out in 300 dpi printer resolution (512/300=1.7). 1024 format produce 3.4 inch (8 x8 cm) at 300 dpi, A 4096 x 4096 image file can produce 32 x 32 cm print-out in 300 dpi. Do you need this size for a single cell if you are not making a poster?. Think about what is the final output size you are going to use, choose a comfortable scan format to work with.

Note about printer resolution: most high quality, professional dye sublimation color printer use 300 dpi. Some journal ask for 600 dpi just for their easy of resizing. Don't using resolution claimed by inkjet printer as the reference for calcuation. 1200 or 2400 dpi might just equal to 300 dpi in dye-sub printer.

Summary:

When choosing scan format, decide which is your priority.
1. Achieving highest optical resolution. Good, everyone likes high resolution image. But how high should it be? Reach the limit of the optical resolution an objective can provide (200 nm or 150 nm that in turn requires 70-100 nm pixel)? Do you need that small or does the specimen provide that small details? Or even if your specimen provide it, after choosing such a small pixel format, is signal intensity and SNR still enough to convey such small detail?
And finally, is it worthwhile to take a painstaking process for a very high pixel-format scan? Will the final image reward your efforts?

2. Meet the requirement of output image size while maintaining good image quality and easy of  image optimization and processing.

3. Compromise between 1 and 2.

For all objective at different excitation wavelength, refer table 2 for their resolution, field size, minimum zoom factor needed for the highest optical resolution of it. Using 1024x1024 scan format and a moderate reduced imaging area (2-3x scanner zoom factor) satisfies most cases.

Table 2.
Optical resolution, magnification, scan field size, zoom factor and digital resolution
see: PDF file

Home |  facilities |  services |  reservation |  confocal tutorial |  other tutorial |  personnel |  link | 
 Statement about this web and tutorial.
For problems or questions regarding this web contact webmaster

This page was last updated 08.12.2009